Sterilant: At 25°C with an immersion time of at least 10 hours.
High Level Disinfectant: At 25°C with an immersion time of at least 45 minutes.
Reuse Period: Glutaraldehyde Solution can be reused for up to a maximum of 14 days, provided the required conditions of glutaraldehyde concentration, pH and temperature exist.
Activation: Activate the Glutaraldehyde Solution by adding the entire contents of the Activator Vial, which is attached to the container. Shake well. Activated solution immediately changes color to green, only indicating that the activator has been added to the solution. Record the date of activation (mixing date) and expiration date on the container label in the space provided, in a logbook or a label affixed to any secondary container used for the activated solution.
Cleaning: Clean thoroughly, rinse and rough dry devices before immersing in Glutaraldehyde Solution. Clean and rinse the lumens of hollow instruments before filling with Glutaraldehyde Solution. Refer to the reusable medical device manufacturer's labeling for additional instructions on disassembly, decontamination, cleaning and leak testing of their equipment.